RJ1836H Multi-function Safety Tester

RJ1836H Multi-functional safety tester, has the test functions of AC voltage withstand, DC voltage withstand, insulation resistance, and it can provide fast and efficient safety testing for battery module, pack, new energy automotive, household appliances, power electronics and other industries.


Multi-function Safety Tester
Main feature
Meet the test requirements of CCC, IEC, EN, VDE, BS, UL, JIS and other mandatory electrical safety standards
Ac withstand voltage, DC withstand voltage, insulation resistance items automatic connection test
Arc detection function, product test arc alarm output lock, prevent continuous arc discharge product burn caused by irreparable damage
Filter function, for capacitive substance test, can relatively maintain data stability in the test, reduce the data surge caused by capacitor charge and discharge
Discharge function, set the discharge time according to the capacitance, display the discharge process voltage, and intuitively understand the discharge result
30 test groups, each group of 8 test steps, easy to operate
RS232 interface, PLC interface, network port communication(LAN)
Technical parameter
Model RJ1836H
AC
Voltage With stand Test
Output Capacity 550 VA (5000V/110mA)
Output Voltage Range/Resolution/ 50V~5000V  /  1V  /  ±(1%rdg.+5dgt.)
Accuracy
Voltage Range/Accuracy 50Hz   /  60Hz   ±0.1%
Frequency
Current   lower Range 0.000mA~110.0mA
limit setting
Current  Upper Range 0.000mA~110.0mA
Limit Setting
Time setting Range/Resolution/ 1.0s ~999.9s /0.1s /±(0.1%rdg.+5dgt.)
Accuracy
Current Range/Resolution/ 0.001mA~110.0 mA  /0.001/ 0.01 /0.1
Measurement Accuracy ±(1%rdg.+5dgt.)
DC
voltage with stand test
Output Capacity 120 VA (6000V/20mA)
Output Voltage Range/Resolution/ 25V~6000V  /  1V  /  ±(1%rdg.+5dgt.)
Accuracy
Current   lower Range 0.0μA~9999μA/10.00mA~20.00mA
limit setting
Current  Upper Range 1.0μA~9999μA/10.00mA~20.00mA
Limit Setting
Time setting Range/Resolution/ 1.0s ~999.9s /0.1s /±(0.1%rdg.+5dgt.)
Accuracy
Current Range/Resolution/ 1.0μA~9999μA/10.00mA~20.00mA    0.1μA/ 1μA /0.01mA   
Measurement Accuracy ±(1%rdg.+5dgt.)
Contact detection current on/off  0.001mA~20.00mA
Maximum Capacitive Load 1μF<1000V,0.75μF<2KV, 0.5uF<3KV,0.08μF < 4KV ,0.04μF < 5KV
Insu
lation resistance test
Output Capacity 120 VA  (6000V/20mA)
Output Voltage Range/Resolution/
Accuracy
25V~6000V  /  1V  /  ±(1%rdg.+5dgt.)
Lower Range 0.100MΩ~100.00GΩ
resistance limit setting
Resistance Range 0.100MΩ~100.00GΩ ;
limit setting HHHH  Indicates that the upper limit is not determined


Time setting Range/Resolution/
Accuracy
1.0s ~999.9s /0.1s /±(0.1%rdg.+5dgt.)
Resistance Range/Accuracy 25V~99V:      
Measurement 0.100MΩ ~500 MΩ      ±(15%rdg.+2dgt.)             

100V~499V: 

0.100MΩ~2.000GΩ      ±(5%rdg.+2dgt.)

500V~799V:  

1.000MΩ~999.9MΩ      ±(2%rdg.+2dgt.)

1.000GΩ~9.999GΩ       ±(5%rdg.+2dgt.)

10.00GΩ~50.00GΩ       ±(15%rdg.+2dgt.)

800V~6000V:     

1.000MΩ~999.9MΩ      ±(2%rdg.+2dgt.)               

1.000GΩ~9.999GΩ       ±(5%rdg.+2dgt.)               

10.00GΩ~100.00GΩ      ±(15%rdg.+2dgt.)
Maximum Capacitive Load 1μF<1000V, 0.75μF<2KV ,0.5μF<3KV ,0.08μF < 4KV,0.04μF < 5KV
Other indicators Arc test 0 off    1~9(1 most sensitive)
Filter function on/off   1~6 (step by step)
External Interfaces RS232(Optional:RS485),PLC,LAN interface
Operating Environment 0℃~40℃,20%RH~75%RH
Power supply interface AC220V,50/60Hz
Net weight 25kg
External Dimension 430mm×133mm×550mm(W×H×D)without foot, foot height 15mm
Attachment 1 manual, 1 power cord (1.5 m), 1 set of communication cable (1.5 m),
1 set of test cable (3 m), 1 D-type connector