RJ684XHigh-speed Safety Comprehensive Tester
RJ684X High-speed Safety Comprehensive Tester is an integrated safety and electrical performance tester with built-in voltage withstand, insulation resistance, grounding resistance, leakage current, power and low-voltage start-up tests. 4 functions performed by the tester can be completed 4 seconds at the fastest, dramatically improving line efficiency.
Main feature
Meet the test requirements of CCC, IEC, EN, VDE, BS, UL, JIS and other mandatory standards for electrical safety
AC voltage withstand, DC voltage withstand, grounding, insulation, leakage, starting, power items automated articulation test;
Has multi-speed power setting, intelligent determination, especially suitable for electric fan, range hood, induction cooker and other multi-speed electrical automatic test
The four safety tests can be completed in 4 seconds at the fastest
30 test groups, 8 test steps per group, convenient operation
Can realize automatic testing
RS232 interface, PLC interface, power communication interface, alarm lamp interface
Technical parameter
Model | RJ6840/41 | ||
Ground resistance test | Open circuit output voltage | <12V | |
Output Current Setting | Range/Resolution/Accuracy | 3A~30A / 1A / ±(3%rdg.+0.18A) | |
Current Frequency Setting | Range/Accuracy | 50Hz / 60Hz / ±0.1% | |
Resistance limit setting | Range/Resolution | 3A~10A: 10mΩ~600mΩ / 1mΩ | |
11A~25A:10mΩ~300mΩ / 1mΩ | |||
26A~30A:10mΩ~200mΩ / 1mΩ | |||
Lower resistance limit setting | Range/Resolution | 0mΩ~100mΩ / 1mΩ | |
Time setting | Range/Resolution/Accuracy | 1.0s~999.9s / 0.1s / ±(0.1%st.+0.5dgt.) | |
Resistance Measurement | Range/Resolution/Accuracy | 1mΩ~600mΩ / 1mΩ / ±(3%rdg.+5dgt.) | |
Insulation Resistance Test | Output Capacity | 2VA(1000V/2mA) | |
Output Voltage | Range/Resolution/Accuracy | 200V~1000V / 1V / ±(3%rdg.+3dgt.) | |
Lower resistance limit setting | Range/Resolution | 0.3MΩ~99.9MΩ/100MΩ~3000MΩ/ 0.1MΩ/1MΩ | |
Resistance limit setting | Range/Resolution | 0.0MΩ~99.9MΩ/100MΩ~3000MΩ/ 0.1MΩ/1MΩ | |
Time setting | Range/Resolution/Accuracy | 1.0s~999.9s / 0.1s / ±(0.1%st.+0.5dgt.) | |
Resistance Measurement | Range/Resolution/Accuracy | 1.0 MΩ~99.9MΩ/0.1MΩ/±(3%rdg.+3dgt.) | |
100 MΩ~2000MΩ/ 1MΩ /±(8%rdg.+8dgt.) | |||
AC Voltage Withstand Test | Output Capacity | 100VA(5000V/20mA) | |
Output Voltage | Range/Resolution/Accuracy | 300V~5000V / 1V / ±(3%rdg.+10dgt.) | |
Voltage Frequency | Range/Accuracy | 50Hz / 60Hz / ±0.1% | |
Current lower limit setting | Range/Resolution | 0.00mA~10.00mA /0.01mA | |
Current Upper Limit Setting | Range/Resolution | 0.10mA~20.00mA /0.01mA | |
Time setting | Range/Resolution/Accuracy | 2.0s~999.9s / 0.1s / ±(0.1%st.+0.5dgt.) | |
Current Measurement | Range/Resolution/Accuracy | 0.10mA~20.00mA/0.01mA/±(3%rdg.+3dgt.) | |
DC voltage withstand test | Produced grain weight | 40VA(4000V/10mA) | |
Output Voltage | Range/Resolution/Accuracy | 300V~4000V / 1V /±(3%rdg.+10dgt.) | |
Current lower limit setting | Range/Resolution | 0.000mA~9.999mA /0.001mA | |
Current Upper Limit Setting | Range/Resolution | 0.010mA~9.999mA /0.001mA | |
Time setting | Range/Resolution/Accuracy | 1.0s~999.9s / 0.1s / ±(0.1%st.+0.5dgt.) | |
Current Measurement | Range/Resolution/Accuracy | 0.001mA~9.999mA/0.001mA/±(3%rdg.+5dgt.) | |
Leakage current test | Current lower limit setting | Range/Resolution | 0μA ~9999μA / 1μA |
Current Upper Limit Setting | Range/Resolution | 0μA ~9999μA / 1μA | |
Time setting | Range/Resolution/Accuracy | 2.0s~999.9s / 0.1s/±(1%st.+2dgt.) | |
Voltage measurement | Range/Resolution/Accuracy | 60V~280V / 1V /±(0.5%rdg.+2dgt.) | |
Current Measurement | Range/Resolution/Accuracy | 50μA ~9999μA / 1μA / | |
DC ~10kHz: ±(3%rdg.+10dgt.) | |||
10kHz~ 1MHz: ±(5%rdg.+50dgt.) | |||
Start-up test | Current upper and lower limit settings | Range/Resolution | 0.00A~25.00A / 0.01A |
Time setting | Range/Resolution/Accuracy | 1.0s~999.9s / 0.1s/±(1%st.+2dgt.) | |
Middle Voltage measurement | Range/Resolution/Accuracy | 60V~280V / 1V /±(0.5%rdg.+2dgt.) | |
Current Measurement | Range/Resolution/Accuracy | 0.030A~3.999A /0.001A/±(0.5%rdg.+20dgt.) | |
4.00A~25.00A /0.01A /±(0.5%rdg.+2dgt.) | |||
Power Test | Power upper and lower limit settings | Range/Resolution | 0W~6000W / 1W |
Time setting | Range/Resolution/Accuracy | 1.0s~999.9s / 0.1s/ ±(1%st.+2dgt.) | |
Voltage measurement | Range/Resolution/Accuracy | 60V~280V / 1V / ±(0.5%rdg.+2dgt.) | |
Current Measurement | Range/Resolution/Accuracy | 0.030A~3.999A /0.001A / ±(0.5%rdg.+20dgt.) | |
4.00A~25.00A /0.01A / ±(0.5%rdg.+2dgt.) | |||
Power Measurement | Range/Resolution/Accuracy | 30.0W~199.9W /0.1W / ±(0.5%rdg.+50dgt.) | |
200.0W~6000.0W /0.1W /±(0.5%rdg.+30dgt.) | |||
Power Factor Measurement | Range/Resolution/Accuracy | 0.10~1.00 /0.01 / ±(2%rdg.+2dgt.) | |
Open-short circuit | Resistance upper and lower limit settings | Range/Resolution | 1Ω~500Ω / 1Ω |
Time setting | Range/Resolution/Accuracy | 1.0s~999.9s / 0.1s/ ±(1%st.+2dgt.) | |
Resistance Measurement | Range/Resolution/Accuracy | 1Ω~500Ω / 1Ω / ±(2%rdg.+2dgt.) | |
Other indicators | External Interfaces | RS232, PLC, power supply communication interface, alarm light interface | |
Operating Environment | 0℃~40℃,20%RH~75%RH | ||
Power Supply | AC220V,50/60Hz | ||
Net weight | 25.3kg | ||
External Dimension | 449mm×151mm×600mm(W×HxD) without foot, foot height 15mm | ||
Attachment | 1 instruction manual, 1 power cable (1.5m), 1 communication cable (1.5m), 1 set of test cable. |