RJ3545Precision DC Resistance Tester

RJ3545 High precision DC resistance tester can measure the resistance of motors, transformers and other coils and the contact resistance of relay contacts through the four-terminal method, and measure the DC resistance of fuses, resistors and various conductive materials. The built-in OVC thermoelectric potential compensation function makes the test results more accurate.

The RJ3545 not only meets the high-precision testing requirements of the laboratory, but also has a variety of communication interfaces to meet the requirements of automatic production line mass testing.


Precision DC Resistance Tester
Main feature
Wide measuring range 0.01μΩ~1200MΩ, 12 steps, max. 6.5 display digits
Maximum test current 1A
Flexible I/O interface, compatible with various test systems
OVC thermoelectric potential compensation function, effectively eliminate the effect of thermoelectric potential
Low current mode test prevents the object under test from fusing, heating or magnetizing
Multi-channel scanning mode can be set, external RJ35-SXX multi-channel scanning unit (optional), to realize multi-channel test
Technical parameter
Model RJ3545
Measuring range 0.01μΩ~1200ΜΩ   12 level
Measuring current 0.1μA~1A
Open Voltage DC 20V (range 10 kΩ~1000MΩ)
Maximum 5.5 V (range 10mΩ~1000Ω)
The maximum DC is 20 mV for LP ON
Measuring Large current mode 10.00000mΩ~1000MΩ
 range Large current OVC pulse mode 10.00000mΩ~1000Ω


Small current mode 100.0000mΩ~100Ω


Large current OVC pulse  mode 100.0000mΩ~100Ω


Low current OVC pulse mode 1000.000mΩ~1000Ω

Measuring Accuracy Large current mode ± (0.06%rdg.+0.08%f.s.)            10mΩ
± (0.06%rdg.+0.01%f.s.)            100mΩ
± (0.012%rdg.+0.008%f.s.)          1000mΩ
± (0.008%rdg.+0.008%f.s.)          10Ω
± (0.007%rdg.+0.001%f.s.)          100Ω
± (0.006%rdg.+0.001%f.s.)          1000Ω
± (0.007%rdg.+0.001%f.s.)          10kΩ
± (0.007%rdg.+0.001%f.s.)          100kΩ
± (0.008%rdg.+0.2%f.s.)            1000kΩ
± (0.03%rdg.+0.2%f.s.)             10MΩ
± (0.2%rdg.+1%f.s.) 100MΩ(100M High-precision mode)
Below 10.00M :±(0.5%rdg.+0.02%f.s.) 100MΩ(100M Low-precision mode)
Above 10.01M :±(1%rdg.+0.02%f.s.) 100MΩ(100M Low-precision mode)
Below 100.0M:±(1%rdg.+0.02%f.s.) 1000MΩ(100M Low-precision mode)
Above 100.1M :±(10%rdg.+0.02%f.s.) 1000MΩ(100M Low-precision mode)
Large current OVC pulse mode ± (0.06%rdg.+0.001%f.s.)           10mΩ
± (0.06%rdg.+0.001%f.s.)           100mΩ
± (0.012%rdg.+0.001%f.s.)          1000mΩ
± (0.008%rdg.+0.005%f.s.)          10Ω
± (0.007%rdg.+0.001%f.s.)          100Ω
± (0.006%rdg.+0.001%f.s.)          1000Ω
Small current mode ± (0.014%rdg.+0.05%f.s.)           100mΩ
± (0.008%rdg.+0.05%f.s.)           1000mΩ
± (0.008%rdg.+0.05%f.s.)           10Ω
± (0.008%rdg.+0.01%f.s.)           100Ω
Large current OVC pulse  mode ± (0.014%rdg.+0.006%f.s.)          100mΩ
± (0.008%rdg.+0.002%f.s.)          1000mΩ
± (0.008%rdg.+0.005%f.s.)          10Ω
± (0.008%rdg.+0.010%f.s.)          100Ω
Low current OVC pulse mode ± (0.2%rdg.+0.003%f.s.)            1000mΩ
± (0.2%rdg.+0.002%f.s.)            10Ω
± (0.2%rdg.+0.002%f.s.)            100Ω
± (0.2%rdg.+0.002%f.s.)            1000Ω
External interface RS232/RS485,communications interface, Handler interface(PLC interface)
Other Functions OVC voltage compensation, comparator function, low current test mode, low current test mode, contact detection function, resistivity test function, temperature test (optional), temperature compensation function (optional)
Usage environment Storage Temperature and humidity range -10°C~ 50°C, 20%RH ~75%RH (no condensation)
Temperature and humidity range 0°C ~40°C, 20%RH~ 75%RH (no condensation)
The temperature and humidity ranges from 20°C~ 26°C and from 20%RH ~75%RH
Power supply AC100V~240V,50/60Hz
Net weight 3.7kg
Dimensions 213mm×88mm×350mm(W×H×D)  without foot, foot height 15mm
Attachment 1 instruction manual, 1 power cable (1.5 m), 1 communication cable (1.5 m), 1 set of test cables (1.5 m),1GPIB connector plug